6 SAMPLE ANALYSIS TOOLS for VOCs, SVOCS, and PESTICIDES

6.2 EX-SITU ANALYSIS

6.2.10 Ion Trap MS


Use: Determines the masses of atoms or molecules found in a solid, liquid, or gas, particularly VOCs.

Description:

Ion trap mass spectrometer uses three electrodes to trap ions in a small volume. The mass analyzer consists of a ring electrode separating two hemispherical electrodes. A mass spectrum is obtained by changing the electrode voltages to eject the ions from the trap. The advantages of the ion trap mass spectrometer include compact size and the ability to trap and accumulate ions to increase the signal-to-noise ratio of a measurement.

Analytes:

1. Non-Halogenated VOCs 2. Non-Halogenated SVOCs 3. Halogenated VOCs
4. Halogenated SVOCs 5. PAHs 6. Pesticides/Herbicides

Media:

Soil/Sediment Water Gas/Air
Requires extraction to liquid or gas phase Requires extraction to liquid or gas phase BETTER
Selectivity: Technique measures the specific contaminant directly.
Susceptibility to Interference: Medium.
Detection Limits : 100-100 ppb (soil); 1-50 ppb (water).
Turnaround Time per Sample: Minutes.
Applicable To:
Screen/Identify Characterize Concentration/Extent Cleanup Performance Long-Term Monitoring
BETTER BETTER BETTER BETTER
Quantitative Data Capability: Produces quantitative data.
Technology Status: Commercially available technology with moderate field experience..
Certification/Verification: Technology has not participated in CalEPA certification and/or CSCT verification program.
Relative Cost per Analysis: Most expensive.

Limitations:

ASTM Standards/EPA Methods:

No applicable ASTM standards or EPA methods are cited for this technology.